MONITORING

4

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

4-69

Figure 4-18 ST Trend Graph

This ST trend graph depicts the changes in STJ from a patient’s first 12-lead ECG over 10 minutes

of monitoring time. The patient’s initial ECG showed no ST elevation in any lead. Then the patient

developed 3 mm elevation in Lead II. This change in ST elevation is represented by the vertical bars

and lasted approximately 5 minutes. (Each vertical bar represents a 30-second interval). After

treatment was initiated, the ST decreased to the current STJ measurement of 1.0, but is still

positive compared to the initial ECG.

The annotation (1.0/1.0) means that the current STJ measurement is elevated 1.0 mm and

represents a change of 1.0 mm from the initial ECG. To confirm the value of the initial 12-lead ECG

STJ measurement, subtract the STJ change from the current STJ measurement, for example,

1.0 – 1.0 = 0. You can display the ST graph of other leads.

Current

STJ

Change in

STJ

Lead

Increase and then
decrease in STJ