Selecting ecg lead, Monitoring the ecg – Physio-Control LIFEPAK 15 User Manual

Page 48

Advertising
background image

4-4

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Monitoring the ECG

Selecting ECG Lead

The LIFEPAK 15 monitor/defibrillator includes two methods for selecting or changing the ECG lead.

To select or change the displayed ECG lead using the

LEAD

button:

:

To select or change the displayed ECG lead using the

SPEED DIAL

:

Note:

The ECG shows dashed lines until the electrodes are connected to the patient.

Note:

When the

VF/VT ALARM

is ON, you are limited to

PADDLES

lead or Lead

II

in Channel 1. See

"Setting Alarms" on page 3-21.

1. Press LEAD. If any ECG lead currently appears

on the Home Screen, the lead changes to
PADDLES

. If PADDLES lead is currently

displayed, the lead changes to Lead II.

2. While the LEAD menu is displayed, press LEAD

again or rotate the SPEED DIAL to the desired
lead.

Note: If lead sets are predefined for Channels 2
and 3, the lead sets show on the menu. The ECG
cable that is connected to the device, such as 3-
lead or 5-wire, determines the leads you can
select. For information about defining lead sets,
see the LIFEPAK 15 Monitor/Defibrillator Setup
Options
provided with your device.

Lead

III

AVR

AVL
AVF

II, CO2

I, II, III

AVR, AVL, AVF

Paddles

I

II

1. For the primary ECG, outline and select

CHANNEL 1

and then select LEAD.

2. Rotate the SPEED DIAL to the desired ECG

lead.

3. Press the SPEED DIAL to select the ECG lead.

4. Repeat this procedure to select or change

displayed ECG waveforms for Channels 2
and 3.

Channel 1

Lead

Size

1.0

Advertising