No shock advised, Automated external defibrillation (aed) – Physio-Control LIFEPAK 15 User Manual

Page 126

Advertising
background image

5-12

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Automated External Defibrillation (AED)

No Shock Advised

The following prompts occur if no shock is advised:

AED Mode

Push ANALYZE

When the CPR countdown time ends, the
PUSH ANALYZE

prompts occur. These prompts

repeat every 20 seconds until you press
ANALYZE

.

AED Mode

No shock advised

If the AED detects a nonshockable rhythm,
the NO SHOCK ADVISED prompts occur. The
defibrillator does not charge, and no shock
can be delivered.

AED Mode

Start CPR

1:46

After NO SHOCK ADVISED, the START CPR
prompts occur. A countdown timer (min:sec
format) continues for the duration specified in
the CPR TIME 2 setup option.

Note: The CPR metronome automatically
provides audible compression “tocks” and
ventilation prompts or tones only during CPR
intervals. To silence the metronome, press
CPR

. To un-silence the metronome, press CPR

again.

AED Mode

Push ANALYZE

When the CPR countdown time ends, the
PUSH ANALYZE

prompts occur. These prompts

repeat every 20 seconds until you press
ANALYZE

.