DATA MANAGEMENT

7

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

7-9

Pacing Event

Figure 7-3 Waveform Event Printout Examples

Memory Capacity

The LIFEPAK 15 monitor/defibrillator retains data for two or more patients when you switch power

off or remove the batteries. The number of patient reports that the LIFEPAK 15 monitor/defibrillator

can store depends on various factors, including the number of displayed waveforms, the duration of

each use, and the type of therapy. The total capacity is 360 minutes of continuous ECG, 90

minutes of continuous data from all channels, or 400 single waveform events. The maximum

memory capacity for a single patient includes up to 200 single waveform reports and 90 minutes of

continuous ECG. When the defibrillator reaches the limits of its memory capacity, the defibrillator

deletes an entire Patient Record using a “first in, first out” priority to accommodate a new Patient

Record. Deleted Patient Records cannot be retrieved.