Technical specifications – HT instruments SOLAR300N User Manual

Page 182

Advertising
background image

SOLAR300 - SOLAR300N

EN - 180

9. TECHNICAL SPECIFICATIONS

9.1. TECHNICAL

FEATURES

Accuracy is given as: [% reading + (number of digits)*resolution] at 23°C

 5°C ; <80%RH


DC Voltage – PV Systems

Range

Accuracy

Resolution Input

impedance

0.0

 1000.0V

(0.5%rdg+2dgt)

0.1V

10M

Voltage values < 20.0V are considered zero by meter

TRMS AC Phase-Neutral/Phase-Earth/Phase-Phase Voltage – Single/ Three-phase PV systems

Range

Accuracy

Resolution Input

impedance

0.0

 600.0V

(0.5%rdg+2dgt)

0.1V

10M

0.0

1000.0V (P-P)

Max crest factor = 2 ; Voltage values < 20.0V are considered zero by meter ; Meter is connectable with VT with ratio from 1

 3000

TRMS DC/AC Phase-Neutral/Phase-Earth/Phase-Phase Voltage – Single/ Three-phase NPV systems

Range

Accuracy

Resolution Input

impedance

0.0

 600.0V

(0.5%rdg+2dgt)

0.1V

10M

0.0

1000.0V (P-P)

Max crest factor = 2 ; Voltage values < 2.0V are considered zero by meter ; Meter is connectable with VT with ratio from 1

 3000

Voltage Anomalies – Phase-Neutral – NPV Single-phase / Three-phase systems with Neutral

Range

Voltage Accuracy

Time (50Hz)

Accuracy

Voltage

resolution

Time (50Hz)

resolution

0.0

 600.0V

(1.0%rdg+2dgt)

10ms

0.2V

10ms

Maximum peak factor = 2. The voltage values < 2.0 V are zeroed.
The instrument can be connected to TV with Factor selectable between 1

 3000

Selectable threshold from

1% to 30%

Voltage Anomalies – Phase-Phase – NPV Three-phase Systems without Neutral

Range

Voltage Accuracy

Time (50Hz)

Accuracy

Voltage

resolution

Time (50Hz)

resolution

0.0

 1000.0V

(1.0%rdg+2dgt)

10ms

0.2V

10ms

Maximum peak factor = 2. The voltage values < 2.0 V are zeroed.
The instrument can be connected to TV with Factor selectable between 1

 3000

Selectable threshold from

1% to 30%

Voltage Spikes – Phase-Ground - Single phase and Three phase systems

Range

Voltage

Accuracy

Voltage

Resolution

Time resolution

(50Hz)

Detection time (50Hz)

-1000V

 100V

(2.0%rdg+60V)

1V

10ms

78

s – 2.5ms

(SLOW)

100V

 1000V

-6000V

 -100V

(10%rdg+100V)

15V

5

s – 160s

(FAST)

100V

 6000V

Voltage detection threshold set from 100V

 5000V

Maximum number of recorded events: 20000

DC Current (with external STD transducer clamp)

Range

Accuracy

Resolution

Input

impedance

Overcharge

protection

0.0

1000.0mV

(0.5%rdg+0.06%FS)

0.1mV

510k

5V

FS = full scale clamp
Measure made by means of Clamp with Output = 1VDC when the clamp is under nominal current.
Current values < 0.1% of FS are zeroed.