HT instruments SOLAR300N User Manual

Page 87

Advertising
background image

SOLAR300 - SOLAR300N

EN - 85

The instrument can displays messages when pressing the F1 key:

Message

Description

Before start a test, please check

zero reading of DC clamp when

it’s not connected to the plant

ALWAYS perform at the start of each test the zeroing of the supplied DC clamp connected to

the meter and not to the PV installation by pressing ZERO key on it. The result should be a
“0.000A” on meter but a small current value up to “0.02A” is permitted. After that, connect the

DC clamp to the installation

Fw SOLAR-01 < 2.02 – Irr.min =

600W/m2

The minimum irradiance threshold different from 600W/m

2

and the internal Fw of SOLAR-01

is < 2.02. The Fw update of SOLAR-01 unit require to send the unit to HT service

Fw SOLAR-01 < 2.03 – IP=2min

The IP > 2min and the internal Fw of SOLAR-01 is < 2.03. In this case the recording is
aborted as the not updated SOLAR-01 can not saving the values in it’s internal memory. The

Fw update of SOLAR-01 unit require to send the unit to HT service

No supply

It is recommended to connect the supply provided with the instrument, before starting
testing/recording. Anyway, it is also possible to perform a testing/recording without supply,

using the instrument’s internal battery.

Wrong phase sequence

The value of parameter “SEQ” in section Measure display (see § 5.2.4)is different from “123”,
which is correct. Check the sequence value of the phases of Voltages V1, V2, V3.

Negative powers

One or more values of the Active Powers P1, P2, P3 are negative (see § 5.2.4). If necessary,

turn the clamps under current by 180° on the conductors in order to obtain always positive

values (except for cases of co-generation).

Solar 01 connected

Solar 02 connected

Confirms the connection between the instrument and SOLAR-0x. This message MUST

ALWAYS appear when connecting the instrument and SOLAR-0x using the USB cable.

Should this message not appear when connecting, disconnect and connect again the USB

cable to the two units.

SOLAR-01 not connected

SOLAR-02 not connected

Check the connection of SOLAR-0x to the instrument. The error message is displayed when

trying to stop a testing or trying to start/stop recording where item SOLAR-0x is flagged,

while no connection to unit SOLAR-0x has been detected.

SOLAR-0x is not recording. Stop

recording.

This message warns the user that a SOLAR-0x unit with no active recording has been

connected to a instrument unit during the testing/recording phase.

Testing in progress

This message is displayed when a testing is being carried out and the GO/STOP key is

pressed. It is not possible to start recording while a test is already being carried out. Stop

testing by pressing the F1 key (FINISH).

Do you want to stop recording?

Message which asks the user for confirmation before stopping the recording/test currently

being carried out.

Wait for recording start

This message warns the user that the recording/test will start when reaching the following

minute or when reaching the date and time set for start (only for recordings on NPV

systems).

Wait for data download

This message warns the user of the data transferring phase from unit SOLAR-0x to the

instrument. Do not disconnect the USB cable during this phase.

Analysis impossible

This message warns the user that the data recorded are not compatible with the set

minimum irradiation threshold (see § 5.3.3.3) and it is not possible to provide any result.

MPP300 connected

This message confirms the connection between the instrument and MPP300. This message
MUST ALWAYS be displayed when connecting the master unit to the MPP300 via USB

cable. If, upon connection, this message should not appear, disconnect the USB cable and

connect it again.

Error in data download from

SOLAR-02

A transmission error has occurred between the SOLAR-02 and SOLAR300N or MPP300

units. Switch off and on again all units and connect them again. The units will try data

transfer again.

MPP300: low battery

The battery of MPP300 is almost flat. Connect the external power supply to MPP300.

No external power supply for

MPP300. Continue?

This message warns that, in order to carry out very long recordings, it is advisable to connect

the external power supply to MPP300.

MPP300: exchanged AC and DC

voltages

An exchange in the connection between DC and AC voltages has been detected. Check the

connections.

MPP300: negative power

One or more Ac negative powers has been measured. This is normally due to a connection
error. Check the connections.

SOLAR-02: not detected. Stop

recording?

You are trying to stop recording while the SOLAR-02 unit is not connected with the MPP300

unit.

SOLAR-02:Firmware not correct

Firmware of SOLAR-02 dosen’t match with SOLAR300N Firmware

MPP300:Firmware not correct

Firmware of MPP300 dosen’t match with SOLAR300N Firmware

Table 17: Description of error before starting recording