SOLAR300 - SOLAR300N

EN - 72

5.4.2.9. Voltage Anomalies
This option permits to set the control parameters relative to voltage anomalies (sags,
swell, supply breaks – see § 10.4) recording which is completely independent from
periodic analysis (regulated by integration period). The below screen is shown by meter:

12/09/2006 – 16:55:10

RECORDING SETTINGS

+ General Parameters
+

Harmonics

Voltage Anom.: 230V 6%-10%

Inrush current: 30A 1s fix

Flicker

Unbalance

132/251 Parameters – Autonomy: 15d 12h

CLP/EXP

PREDEF.

SEL

MODIFY

Fig. 127: Voltage Anomalies selection screen

Press F4 key (alternatively touch MODIFYat display) for setting the below voltage
anomalies parameters:

Reference nominal voltage Vref depending on the type of considered system. In

particular Vref = VP-N (Single phase and three phase 4-wire systems), Vref = VP-P
(three phase 3-wire and ARON systems).

The higher threshold percentage of reference nominal voltage, selectable from 1% to

30% for voltage swell detection.

The lower threshold percentage of reference nominal voltage, selectable from 1% to

30% for voltage sags or breaks detection.

12/09/2006 – 16:55:10

RECORDING SETTINGS

+ General Parameters
+

Harmonics

Voltage Anom.: 230V 6%-10%

Inrush current: 30A 1s fix

Flicker

Unbalance

Voltage Anom.:

230V

6% 10%

MOD(+)

MOD(-)

Fig. 128: Voltage Anomalies screen - setting control parameters

1. Move the cursor using arrows keys on field relative to “Voltage Anom.” , marked it with

blue background.

2. Use up or down arrow keys or press F3 or F4 keys (alternatively touch the MOD(+) or

MOD(-) s at display) to setting the correspondent value. Pressing and holding of these
keys permits a rapid values setting, while the single pressure increase or decrease on
only unit the value. Use left or right arrow keys to move on various fields.

3. Press SAVE or ENTER keys (or the smart icon

) to save each performed setting.

CAUTION

The nominal reference value should be set depending on the type of
considered system on test and the measured voltage. A message “Wrong
Vref voltage anomalies
” at the recording start is shown by meter which do
not permits the recording operation due to this wrong configuration (e.g.: 4-
wire system and Vref = 400V). Set the correct value in this case.