Measure:voltage:ac, Measure:voltage:dc:ratio:digit – RIGOL M300 User Manual

Page 22

Advertising
background image

RIGOL

Chapter 2 Command System

Programming Guide for DM3000

2-10

Function

The query returns the ratio of two DC voltages in scientific notation, for
example, +1.74214858E-01.

Explanation The instrument should input two DC voltages at the same time.

9. :MEASure:VOLTage:DC:RATIo:DIGIt
Syntax

:MEASure:VOLTage:DC:RATIo:DIGIt?
:MEASure:VOLTage:DC:RATIo:DIGIt {INC|DEC|5|6|7}

Function

Set the display digit for the ratio of two DC voltages.
The query returns 5, 6 or 7.

Explanation Refer to the “Explanation” in :MEASure:VOLTage:DC:DIGIt.
Example

Set the ratio display digit to 6:
:MEASure:VOLTage:DC:RATIo:DIGIt 6
Decrease the display digit by one:
:MEASure:VOLTage:DC:RATIo:DIGIt DEC


10. :MEASure:VOLTage:AC?
Syntax

:MEASure:VOLTage:AC?

Function

The query returns the AC voltage measurement value in the scientific
notation (for example, +6.59000527E-03) and the unit is V.


11. :MEASure:VOLTage:AC
Syntax

:MEASure:VOLTage:AC {0|1|2|3|4|MIN|MAX|DEF}

Function

Set the range of AC voltage measurement.

Explanation The range corresponding to each parameter is as follows.

Parameter

Range

0

200 mV

1

2 V

2

20 V

3

200 V

4

750 V

MIN

200 mV

MAX

750 V

DEF

20 V

Example

Set the range to the minimum:
:MEASure:VOLTage:AC 0 or
:MEASure:VOLTage:AC MIN