Chapter 2 Command System

Programming Guide for DM3000

2-12

17. :MEASure:VOLTage:AC:FREQuency:HIDE
Syntax

:MEASure:VOLTage:AC:FREQuency:HIDE

Function

Hide the frequency on the secondary screen during AC voltage
measurement.

Explanation The command is valid only when the frequency display on the

secondary display is enabled in AC voltage measurement.


18. :MEASure:VOLTage:AC:FREQuency:STATe?
Syntax

:MEASure:VOLTage:AC:FREQuency:STATe?

Function

Query whether the frequency is displayed on the secondary screen
during AC voltage measurement.
The query returns DISPLAY or HIDE.


19. :MEASure:CURRent:DC?
Syntax

:MEASure:CURRent:DC?

Function

The query returns the DC current measurement value in scientific
notation (for example, -3.74725404E-06) and the unit is A.


20. :MEASure:CURRent:DC
Syntax

:MEASure:CURRent:DC {0|1|2|3|4|MIN|MAX|DEF}

Function

Set the range of DC current measurement.

Explanation  The measurement mode will change to “Manual” while you set the

range.

 The range corresponding to each parameter is as follows.

Parameter

Range

0

2 mA

1

20 mA

2

200 mA

3

1 A

4

10 A

MIN

2 mA

MAX

10 A

DEF

200 mA

Example

Set the range of DC current measurement to the maximum:
:MEASure:CURRent:DC 4 or
:MEASure:CURRent:DC MAX