Chapter 2 Command System

Programming Guide for DM3000

2-36

1. :TRIGger:SOURce
Syntax

:TRIGger:SOURce?
:TRIGger:SOURce {AUTO|SINGLE|EXT}

Function

Set the trigger source of the measurement to AUTO, SINGLE or
EXT.
The query returns AUTO, SINGLE or EXT.

Default

AUTO


2. :TRIGger:AUTO:INTErval
Syntax

:TRIGger:AUTO:INTErval?
:TRIGger:AUTO:INTErval

<value>

Function

Set the display interval of the multimeter. The default unit is ms.
The query returns the interval with the unit (ms).

Explanation

The range of

<value>

is related to the current reading resolution.

DM306x

Reading resolution

<value>

Default

4 ½

(ACV/ACI, 3 ½)

30 ms - 7000 ms

30 ms

5 ½

200 ms - 7000 ms

200 ms

ACV/ACI, 4 ½

300 ms - 7000 ms

300 ms

6 ½

(ACV/ACI, 5 ½)

400 ms - 7000 ms

400 ms

DM305x

Reading resolution

<value>

Default

3 ¾

(ACV/ACI/DCI, 3 ½)

30 ms - 7000 ms

30 ms

4 ¾

(DCI, 4 ½)

200 ms - 7000 ms

200 ms

ACV/ACI, 4 ½

300 ms - 7000 ms

300 ms

5 ¾

(ACV/ACI/DCI, 5 ½)

400 ms - 7000 ms

400 ms

Example

Set the interval to 1000 ms:
:TRIGger:AUTO:INTErval 1000
The query returns 1000 ms.


3. :TRIGger:AUTO:HOLD
Syntax

:TRIGger:AUTO:HOLD?