Chapter 2 Command System

RIGOL

Programming Guide for DM3000

2-13

21. :MEASure:CURRent:DC:RANGe?
Syntax

:MEASure:CURRent:DC:RANGe?

Function

Query the range of DC current measurement.
The query returns 0, 1, 2, 3 or 4.


22. :MEASure:CURRent:DC:DIGIt
Syntax

:MEASure:CURRent:DC:DIGIt?
:MEASure:CURRent:DC:DIGIt {INC|DEC|5|6|7}

Function

Set the display digit of DC current measurement value.
The query returns 5, 6 or 7.

Explanation Refer to the “Explanation” in :MEASure:VOLTage:DC:DIGIt.
Example

Set the display digit of DC current measurement value to 6:
:MEASure:CURRent:DC:DIGIt 6
Decrease the display digit by one:
:MEASure:CURRent:DC:DIGIt DEC


23. :MEASure:CURRent:AC?
Syntax

:MEASure:CURRent:AC?

Function

The query returns the AC current measurement value in scientific
notation (for example, +4.29493009E-05) and the unit is A.


24. :MEASure:CURRent:AC
Syntax

:MEASure:CURRent:AC {0|1|2|3|MIN|MAX|DEF}

Function

Set the range of AC current measurement.

Explanation  The measurement mode will change to “Manual” while you set the

range.

 The range corresponding to each parameter is as follows.

Parameter

Range

0

20 mA

1

200 mA

2

2 A

3

10 A

MIN

20 mA

MAX

10 A

DEF

200 mA

Example

Set the range of AC current measurement to the maximum:
:MEASure:CURRent:AC 3 or