Parameter

Range

DM306x

DM305x

0

2 nF

4 nF

1

20 nF

40 nF

2

200 nF

400 nF

3

2 uF

4 uF

4

20 uF

40 uF

5

200 uF

200 uF

MIN

2 nF

4 nF

MAX

200 uF

200 uF

DEF

200 nF

400 nF

Example

Set the capacitance range to the maximum:
:MEASure:CAPacitance 5 or
:MEASure:CAPacitance MAX


53. :MEASure:CAPacitance:RANGe?
Syntax

:MEASure:CAPacitance:RANGe?

Function

Query the capacitance measurement range.
The query returns 0, 1, 2, 3, 4 or 5.


54. :MEASure:CAPacitance:DIGIt
Syntax

:MEASure:CAPacitance:DIGIt?
:MEASure:CAPacitance:DIGIt {INC|DEC|5|6|7}

Function

Set the display digit of capacitance measurement value.
The query returns 5, 6 or 7.

Explanation Refer to the “Explanation” in :MEASure:VOLTage:DC:DIGIt.
Example

Set the display digit of capacitance measurement value to 6:
:MEASure:CAPacitance:DIGIt 6
Decrease the display digit by one:
:MEASure:CAPacitance:DIGIt DEC