Measure:current:ac:frequency:state, Measure:resistance, Measure:resistance:range – RIGOL M300 User Manual

Page 27

Advertising
background image

Chapter 2 Command System

RIGOL

Programming Guide for DM3000

2-15

secondary display is enabled in AC current measurement.


30. :MEASure:CURRent:AC:FREQuency:STATe?
Syntax

:MEASure:CURRent:AC:FREQuency:STATe?

Function

Query whether the frequency is displayed on the secondary screen
during AC current measurement.
The query returns DISPLAY or HIDE.


31. :MEASure:RESistance?
Syntax

:MEASure:RESistance?

Function

The query returns the 2-wire resistance measurement value in
scientific notation and the unit is Ω.


32. :MEASure:RESistance
Syntax

:MEASure:RESistance {0|1|2|3|4|5|6|MIN|MAX|DEF}

Function

Set the range of 2-wire resistance measurement.

Explanation  The measurement mode will change to “Manual” while you set the

range.

 The range corresponding to each parameter is as follows.

Parameter

Range

DM306x

DM305x

0

200 Ω

400 Ω

1

2 kΩ

4 kΩ

2

20 kΩ

40 kΩ

3

200 kΩ

400 kΩ

4

1 MΩ

4 MΩ

5

10 MΩ

100 MΩ

6

100 MΩ

MAX

100 MΩ

100 MΩ

MIN

200 Ω

400 Ω

DEF

200 kΩ

400 kΩ

Example

Set the range of 2-wire resistance measurement to the minimum:
:MEASure:RESistance 0 or
:MEASure:RESistance MIN


33. :MEASure:RESistance:RANGe?
Syntax

:MEASure:RESistance:RANGe?