Chapter 2 Command System

RIGOL

Programming Guide for DM3000

2-11

12. :MEASure:VOLTage:AC:RANGe?
Syntax

:MEASure:VOLTage:AC:RANGe?

Function

Query the range of AC voltage measurement.
The query returns 0, 1, 2, 3 or 4.


13. :MEASure:VOLTage:AC:FILTer
Syntax

:MEASure:VOLTage:AC:FILTer?
:MEASure:VOLTage:AC:FILTer {SLOW|MID|FAST}

Function

Set the AC filter mode to Slow, Mid or Fast.
The query returns SLOW, MID or FAST.

Default

FAST


14. :MEASure:VOLTage:AC:DIGIt
Syntax

:MEASure:VOLTage:AC:DIGIt?
:MEASure:VOLTage:AC:DIGIt {INC|DEC|5|6|7}

Function

Set the display digit of AC voltage measurement value.
The query returns 5, 6 or 7.

Explanation Refer to the “Explanation” in :MEASure:VOLTage:DC:DIGIt.
Example

Set the display digit of AC voltage measurement value to 6:
:MEASure:VOLTage:AC:DIGIt 6
Decrease the display digit by one:
:MEASure:VOLTage:AC:DIGIt DEC


15. :MEASure:VOLTage:AC:FREQuency?
Syntax

:MEASure:VOLTage:AC:FREQuency?

Function

The query returns the current frequency of AC voltage measurement in
scientific notation (for example, +5.30803456e+02) and the unit is Hz.

Explanation The multimeter should work under the AC voltage measurement mode

when you use this command.


16. :MEASure:VOLTage:AC:FREQuency:DISPlay
Syntax

:MEASure:VOLTage:AC:FREQuency:DISPlay

Function

Display the frequency of the AC signal on the secondary screen during
AC voltage measurement.

Explanation The multimeter should work under the AC voltage measurement mode

when you use this command.