Chapter 2 Command System

Programming Guide for DM3000

2-38

The query returns RISE or FALL.

Default

RISE


8. :TRIGger:VMComplete:POLAr
Syntax

:TRIGger:VMComplete:POLAr?
:TRIGger:VMComplete:POLAr {POSitive|NEGative}

Function

Set the polarity of the VMC output at the rear panel.
The query returns POSITIVE or NEGATIVE.

Default

POSitive


9. :TRIGger:VMComplete:PULSewidth
Syntax

:TRIGger:VMComplete:PULSewidth?
:TRIGger:VMComplete:PULSewidth

<value>

Function

Set the pulse width of the VMC output at the rear panel. The default
unit is ms.
The query returns the pulse width.

Explanation

The range of

<value>

is related to the current reading resolution.

DM306x

Reading resolution

<value>

4 ½

(ACV/ACI, 3 ½)

1 ms - 29 ms

5 ½ (ACV/ACI, 4 ½)

1 ms - 199 ms

6 ½ (ACV/ACI, 5 ½)

1 ms - 399 ms

DM305x

Reading resolution

<value>

3 ¾

(ACV/ACI/DCI, 3 ½)

1 ms - 29 ms

4 ¾

(ACV/ACI/DCI, 4 ½)

1 ms - 199 ms

5 ¾

(ACV/ACI/DCI, 5 ½)

1 ms - 399 ms

Example

Set the pluse width to 100 ms:
:TRIGger:VMComplete:PULSewidth 100
The query returns 100.