Resolution:current:dc – RIGOL M300 User Manual

Page 35

Advertising
background image

Chapter 2 Command System

RIGOL

Programming Guide for DM3000

2-23

:RESOlution:VOLTage:AC {0|1|2|MIN|MAX|DEF}

Function

Set the reading resolution of AC voltage measurement.
The query returns 0, 1 or 2.

Explanation

 The AC voltage measurement function must be enabled before

using this command.

 The “DEF” is 1.
 The reading resolution corresponding to each parameter is as

follows.

Range

Reading resolution

0

3 ½ digits

1

4 ½ digits

2

5 ½ digits

MAX

5 ½ digits

MIN

3 ½ digits

DEF

4 ½ digits

Example

Set the reading resolution of AC voltage measurement to 5 ½
digits:
:RESOlution:VOLTage:AC 2


4. :RESOlution:CURRent:DC
Syntax

:RESOlution:CURRent:DC?
:RESOlution:CURRent:DC {0|1|2|MIN|MAX|DEF}

Function

Set the reading resolution of DC current measurement.
The query returns 0, 1 or 2.

Explanation

 The DC current measurement function must be enabled before

using this command.

 The reading resolution of each parameter.

Range

Reading resolution

DM306x

DM305x

0

4 ½ digits

3 ½ digits

1

5 ½ digits

4 ½ digits

2

6 ½ digits

5 ½ digits

MAX

6 ½ digits

5 ½ digits

MIN

4 ½ digits

3 ½ digits

DEF

5 ½ digits

4 ½ digits

Example

Set the reading resolution of DC current measurement to 5 ½
digits:

Advertising