<Appendix 1. List of Parameters for Each Block of the YTA>

A1-9

IM 01C50T02-01E

Relative

Index

Index

Parameter Name

Default

(factory setting)

Write

Description

36

2036 SENSOR_RANGE_

2

Range of sensor

Range of sensor 2.

37

2037 SENSOR_SN_2

Auto

Serial number of sensor 2.

38

2038 SENSOR_CAL_

METHOD_2

103

Auto

Calibration method for sensor 2:

103 = Factory trim standard calibration

104 = User trim standard calibration

39

2039 SENSOR_CAL_

LOC _2

Auto

Shows and is used to record the location

where sensor 2 was calibrated.

40

2040 SENSOR_CAL

_DATE_2

Auto

Shows and is used to record the date when

sensor 2 was calibrated.

41

2041 SENSOR_CAL

_WHO_2

Auto

Shows and is used to record the person who

calibrated sensor 2.

42

2042 SENSOR _

CONNECTION_2

As specified by the

customer before

shipment

O/S

Number of connection wires of sensor 2.

43

2043 SECONDARY

_VALUE

Indicates the terminal board temperature.

44

2044 SECONDARY

VALUE_UNIT

°C

Auto

Unit of the terminal board temperature.

45

2045 MODULE_SN

Serial number

Serial number

46

2046 ALARM_SUM

Enable

Shows the alarm summary for the transducer

block.

47

2047 PRIMARY_VALUE

_FTIME_1

2 second

O/S

Time constant (in seconds) of the first-order

lag filter applied to the sensor 1 input . (0 to

99)

48

2048 CAL_STATE_1

0 (User Cal Off)

O/S

Indicates the validity of user calibration for

sensor 1:

0 =User Cal Off(Invalidate user-set

calibration values)

1 = User Cal On(Validate user-set calibration

values)

2 = Calibration Exec(User calibration mode)

49

2049 CJC_SELECT_1

0 (internal CJC)

O/S

Selects whether the terminal board

temperature or user-set constant

(CONSTANT_CJC_TEMP_1) is to be used

for cold junction compensation (CJC) for the

sensor 1 input. Valid for Thermocouple input

only.

0=Internal CJC

1=Constant CJC

50

2050 CONSTANT_CJC

_TEMP_1

O/S

User-set constant for CJC for the sensor 1

input. Setting 0 in this parameter disables

RJC. Valid only when CJC_SELECT_1 is set

to 1.

51

2051 WIRING _

RESISTANCE_1

0

O/S

Wiring resistance of the sensor 1 input. For a

2-wire resistance input, the input resistance

minus this value is used as the temperature

value.

52

2052 SENSOR_MATCH

_R0_1

0

O/S

Value of the factor R0 used in the sensor

matching function for the sensor 1 input

53

2053 SENSOR_MATCH

_A_1

0

O/S

Value of the factor A used in the sensor

matching function for the sensor 1 input

54

2054 SENSOR_MATCH

_B_1

0

O/S

Value of the factor B used in the sensor

matching function for the sensor 1 input

55

2055 SENSOR_MATCH

_C_1

0

O/S

Value of the factor C used in the sensor

matching function for the sensor 1 input

56

2056 SENSOR_MATCH

_ALPHA_1

0

O/S

Value of the factor α used in the sensor

matching function for the sensor 1 input

57

2057 SENSOR_MATCH

_DELTA_1

0

O/S

Value of the factor δ used in the sensor

matching function for the sensor 1 input

Advertising