Fluke Biomedical ProSim 6 User Manual

Page 101

Advertising
background image

Glossary

Introduction

A

A-7

QRS Complex

The part of the P-QRS-T wave that records ventricular depolarization and contraction.

R-Value

The non-pulsating components of tissue, specifically the tissue bed, the venous blood, the

capillary blood, and nonpulsatile arterial blood. Also referred to as the DC component.

Resistance

The opposition to electric current that is characteristic of a medium, substance, or circuit

element.

SA Node

The dominant pacemaker site in the heart, responsible for setting the heart rate.

Positioned in the right atrium near the inlet of the superior vena cava.

SaO2

Abbreviation for saturated oxygen, SaO2 is the ratio of the concentration of

oxyhemoglobin (cHbO2) to the concentration of the two principle types of blood

hemoglobin: saturated hemoglobin (HbO2) plus reduced hemoglobin (Hb).

Serial Port

An asynchronous COMmunication port/address to which a peripheral—such as a printer

or a mouse—is connected to a computer or other device. SEE RS-232.

Sinus Arrhythmia

Beats that are normal, but triggered at an irregular rate, from 60 BPM to 100 BPM.
Sinus arrhythmia occurs when the SA node paces the heart irregularly. Typically, the

heartbeat increases with each intake of breath and decreases with each exhalation (a

condition most commonly found in young children and the elderly).

SpO2

The type of saturated oxygen measured with a pulse oximeter.

Supraventricular Tachycardia

Normal rhythm at a faster-than-normal rate of 200 BPM. Supraventricular tachycardia is

a combination of a junctional tachycardia (that is, an atrial tachycardia occurring in the

AV or junctional node) and an atrial tachycardia. Therefore, supraventricular tachycardia

encompasses multifocal, ectopic, atrial pacemakers in and around the AV node above the

bundle of His.

Swan-Ganz

A soft, balloon-tipped catheter used for measuring blood pressure and cardiac output.

The catheter is guided by blood flow into the pulmonary artery. A monitor near the tip of

the catheter detects PAP, PCWP, and thermodilution.

Swan-Ganz Procedure

After insertion of a Swanz-Ganz catheter into a vein (usually the basilic vein of the

forearm), the catheter is gently guided by the flow of the blood into the pulmonary artery.

A monitor attached to the distal lumen port supplies a reading of pulmonary-artery

pressure (PAP). Pulmonary-capillary-wedge pressure (PCWP) is determined by inflating

the balloon, which becomes wedged; when this wedge blocks blood flow, it provides a

reading of the pressure in the left side of the heart.

Thermodilution

The measuring of temperature change, enabled by the injection of a cold or room-

temperature solution (such as saline) into the right atrium by means of a Swan-Ganz

procedure.

Advertising
This manual is related to the following products: