Invasive blood pressure simulation and tests, How to set the invasive blood pressure variables – Fluke Biomedical ProSim 6 User Manual

Page 41

Advertising
background image

Vital Signs Simulator

Invasive Blood Pressure Simulation and Tests

29

glh027.bmp

Figure 17. Fetal ECG Intrauterine Pressure Screen

If the contraction value is set to Manual Start, only one contraction cycle is simulated.
The timer shows the time until the contraction ends and the subsequent contraction starts.
Push the Stop softkey to stop contractions and go back to the Fetal ECG: IUP screen.

Invasive Blood Pressure Simulation and Tests

The Product simulates blood pressure for Invasive blood pressure monitors. Each blood

pressure variable can be set through the front-panel controls. The Product also

simulations Swan-Ganz and cardiac catheterization procedures.

How to Set the Invasive Blood Pressure Variables

The Product can simulate two invasive blood pressure transducers at one time. The blood

pressure variables are set separately for each channel. To set these variables, push 

to show the IBP screen in Figure 18 in the display.

glh009.bmp

Figure 18. Invasive Blood Pressure Screen

To set the chamber:
1. Push  or  to highlight the Chamber value on channel 1 or channel 2.
2. Push .
3. Push  or  to highlight a value in the chamber list. Below is a list of chamber

values in the Product.

Static Right

Ventricle

Arterial Pulmonary

Artery

Radial Artery

PA Wedge

Left Ventricle

Right Atrium (CVP)

Left Atrium