How to turn the product on, How to connect a pc to the product – Fluke Biomedical ProSim 6 User Manual

Page 19

Advertising
background image

Vital Signs Simulator

How to Turn the Product On

7

How to Turn the Product On

Push  on the front panel to turn the Product on. The screen shown in Figure 3 shows

the power-up screen.

glh002.bmp

Figure 3. Power-Up Screen

When the self test is complete and no errors are sensed, the home screen in Figure 4

shows in the display.

glh001.bmp

Figure 4. Home Screen

From the home screen, a number of pre-defined patient simulations can be used to set all

the simulation functions at one time. See Pre-Defined Simulations to learn more on how

to use simulations. Waveform and ECG rate parameters are set through the ECG

function. See the ECG Function section to learn more on how to set these variables. The

respiration rate (Resp Rate), temperature parameter, cardiac output, and fetal simulation

are set through the special functions menu. The two Invasive Blood Pressure (IBP)

channels are set through the IBP function. The Non-Invasive Blood Pressure variables are

set through the NIBP function. See the Blood Pressure Simulations and Tests section to

learn more on how to set IBP and NIBP variables. The SpO2 variable is set through the

SpO2 function. See the Oximeter SpO2 Optical Emitter and Detector section to learn

more on how to set this variable.

How to Connect a PC to the Product

You can use a PC to store presets, auto sequences, and test results from the Product. You

must use the ProSim Mini Plug-In on the PC to communicate with the Product.

Note

You must have Ansur Test Executive version 2.9.6 on your PC to
communicate with the Product.