How to set the spo2 parameters – Fluke Biomedical ProSim 6 User Manual

Page 67

Advertising
background image

Vital Signs Simulator

Oximeter SpO2 Optical Emitter and Detector

55

ProSim 8

SpO

2

Finger Module

glh045.eps

Figure 49. Magnetic Holder for SpO2 Artificial Finger

How to Set the SpO2 Parameters

You can raise or lower the degree of oxygen saturation. To change the SpO2 value:
1. Push  or  to highlight the SpO2 value.
2. Push .
3. Push  or  to increase or decrease the SpO2 value in 1 % steps between 30 and

100 %.

4. Push  to set that SpO2 value and go back to the SpO2 screen.
To change the pulse amplitude value:
1. Push  or  to highlight the Pulse Amplitude value.
2. Push .
3. Push  or  to change the pulse amplitude value between 0 and 20 %. Each push of

a key changes the value 0.01 % between 0 and 1 % in the direction of the key pushed.

The value changes 0.1 % between 1 and 10 %. The value changes 1 % between 10

and 20 %.

4. Push  to set the pulse amplitude value and go back to the SpO2 screen.
To change the transmission value:
1. From the SpO2 screen, push  or  to highlight the Transmission value.
2. Push .
3. Push  or  to highlight Dark, Thick Finger, Medium Finger, Light, Thin

Finger, and Neonatal Foot.

4. Push  to set that transmission value and go back to the SpO2 screen.