Fluke Biomedical ProSim 6 User Manual

Page 68

Advertising
background image

ProSim 6/8
Users Manual

56

You can change the simulated heart rate of the Product from the SpO2 screen. To change

the heart rate:
1. Push  or  to highlight the Heart Rate value.
2. Push .
3. Push  or  to adjust the heart rate. Each push of a key moves the heart rate one

beat in the direction of the key pushed. The range is 10 to 300 beats/minute (BPM).

4. Push  to go back to the SpO2 screen.
To change SpO2 respiration:
1. From the SpO2 screen, push the  or  to highlight the Respiration value.
2. Push .
3. Push  or  to highlight 1%, 2%, 3%, 4%, 5% or off.
4. Push  to set the respiration and go back to the SpO2 screen.
You can test SpO2 under different ambient light conditions. To change the ambient light

value:
1. From the SpO2 screen, push the  or  to highlight the Ambient Light value.
2. Push .
3. Push  or  to highlight an ambient light value. Below is the list of ambient light

values in the Product.

Sunlight

Indoor – 4 kHz

Indoor – DC

Indoor – 5 kHz

Indoor – 50 Hz

Indoor – 6 kHz

Indoor – 60 Hz

Indoor – 7 kHz

Indoor – 1 kHz

Indoor – 8 kHz

Indoor – 2 kHz

Indoor – 9 kHz

Indoor – 3 kHz

Indoor – 10 kHz

4. Push  to go back to the SpO2 screen with the new ambient light value.
The manufacturer must be known before you do a pulse oximeter test, optically through

an artificial finger. For Masimo, you need to know if the sensor is a 2 wavelength or

Rainbow sensor. You can configure the Product for the make of pulse oximeter(s) used

for the test. You change variables for each of the pulse oximeters through the type

variable. To change the oximeter sensor type:
1. From the SpO2 screen, push the  or  to highlight the Type value.
2. Push .