How to do an nibp monitor test – Fluke Biomedical ProSim 6 User Manual

Page 57

Advertising
background image

Vital Signs Simulator

Non-Invasive Blood Pressure Simulation and Tests

45

glh005.bmp

Figure 35. Blood Pressure Measurement Graph

Note

The Summary and Graph softkeys show in the NIBP screen after you do an
NIBP simulation.

To see the measured parameters of the last blood pressure measurement cycle, push the

Summary softkey in the NIBP screen. To see the graph of the last NIBP simulation,

push the Graph softkey in the NIBP screen,

How to Do an NIBP Monitor Test

To do an accuracy test on an NIBP monitor:
1. Connect the NIBP monitor to the Product as shown in Figure 37.
2. Start an NIBP pressure cycle on the monitor. Refer to the monitor manual as

necessary. After you start the blood pressure measurement cycle:

The blood pressure cuff inflates around the mandrel.

The Product starts the peripheral pulse simulation and shows the blood pressure

measurement graph in the display.

Note

The graph shows in the display automatically only when the blood pressure
measurement cycle starts when the NIBP screen shows in the display.

The Product starts blood pressure simulation when the pressure is 10 mmHg.

Heart beat simulation starts when the pressure equals the diastolic pressure set

into the Product.

The Product stops the simulation when the pressure is 2 mmHg or below.

The NIBP monitor interprets and shows the measured blood pressure values and

heart rate when the test stops.

3. Push the Summary softkey to show the blood pressure measurements in Figure 36

in the display.

Advertising
This manual is related to the following products: