How to simulate intrauterine pressure (iup) – Fluke Biomedical ProSim 6 User Manual

Page 38

Advertising
background image

ProSim 6/8
Users Manual

26

glh025.bmp

Figure 14. Fetal ECG Screen

Push the softkey labeled Back to go back to the Special Functions screen.
To change the fetal heart rate:
1. Push  or  to highlight the Fetal HR value.
2. Push .
3. Push  or  to change the heart rate value in 1 bpm steps between 60 and 240 bpm.

Note

When you hold down the direction key, the step size will change to 10 bpm
until the key is released.

4. Push  to set the heart rate and go back to the Fetal ECG screen.
To change the amplitude of the fetal waveform:
1. Push  or  to highlight the Amplitude value.
2. Push  or  to adjust the amplitude. Each push of a key increases or decreases the

amplitude in the direction of the key pushed. The Product simulates a fetal heart

signal amplitude of 0.05 to 0.25 mV by 0.05 mV intervals and 0.50 to 5.00 mV that is

set in 0.25 mV intervals.

3. Push  to set the amplitude and go back to the Fetal ECG screen.

How to Simulate Intrauterine Pressure (IUP)

The simulated intrauterine-pressure (IUP) waveform shows a measurement read by an

intra-amniotic catheter connected to a pressure transducer. The Product simulates the IUP

on IBP channel 1, with a 5 or 40

μ

V/V/mmHg sensitivity (as has been setup for blood

pressure).
The Product sends waveforms to simulate intrauterine pressure during a contraction of

the uterus in childbirth. Each IUP wave goes for 90 seconds with a bellshaped pressure

curve that increases from zero to 90 mmHg and goes back to zero. In an IUP-wave

simulation, the fetal heart rate (which always begins at 140 BPM, regardless of the fetal

ECG rate.) changes with the blood pressure. The fetal heart rate and blood pressure

shows in the display.
The IUP period is adjustable to four preconfigured values: a manually started contraction,

and contractions that start at 2, 3, or 5 minute intervals.
To do an Intrauterine pressure simulation, connect the fetal monitor to the Product as

shown in Figure 15.

Advertising