How to set the cardiac output waveform, How to set the baseline temperature, How to set injectate temperature – Fluke Biomedical ProSim 6 User Manual

Page 53: How to start a cardiac output simulation, Non-invasive blood pressure simulation and tests

Advertising
background image

Vital Signs Simulator

Non-Invasive Blood Pressure Simulation and Tests

41

How to Set the Cardiac Output Waveform

To set the cardiac output waveform:
1. In the Cardiac Output screen, use  or  to highlight the wave parameter.
2. Push .
3. Push  or  to highlight a waveform name in the list of waveforms.
Table 13 shows the cardiac output waveforms for the Product.

Table 13. Cardiac Output Waveforms

Waveform Description

2.5 L/min

Normal waveform with accuracy of 2.5 L/min

5.0 L/min

Normal waveform with accuracy of 5.0 L/min

10.0 L/min

Normal waveform with accuracy of 10.0 L/min

Interrupted Injectable

Interrupted injection waveform

LR Shunt

Left to right ventricular shunt

Calibrated Pulse

Calibrated square wave pulse

4. Push  to set the wave and go back to the Cardiac Output screen.

How to Set the Baseline Temperature

To set the baseline temperature:
1. In the cardiac output screen, if not already highlighted, use  or  to highlight the

baseline temperature value.

2. Push .
3. Push  or  to highlight 36

°

C, 37

°

C, or 38

°

C.

4. Push  to set the baseline temperature and go back to the Cardiac Output

screen.

How to Set Injectate Temperature

To set the injectate temperature:
1. In the cardiac output screen, use  or  to highlight the injectate temperature value.
2. Push .
3. Push  or  to highlight 0

°

C or 24

°

C.

4. Push  to set the temperature and go back to the Cardiac Output screen.
As you change the injectate temperature, the calibration coefficient necessary for the

monitor is shown in the display.

How to Start a Cardiac Output Simulation

After you set the parameters for a cardiac output simulation, push the Start softkey. The

simulation completes automatically. To stop the simulation, push the Stop softkey.

Non-Invasive Blood Pressure Simulation and Tests

The Product simulates blood pressure for Non-Invasive blood pressure monitors. Each

blood pressure variable can be set through the front-panel controls. The Product also does

leak, pressure source, and pressure relief tests. The manometer function sets the Product

to measure static pressure and shows the pressure in the display.