How to simulate respiration – Fluke Biomedical ProSim 6 User Manual

Page 48

Advertising
background image

ProSim 6/8
Users Manual

36

To set the simulated temperature:
1. Push .
2. Push , , , or  to highlight the Temperature selection then push  to

show the set temperature in the display.

3. Push .
4. Push  or  to change the temperature in 0.5

°

C steps between 30.0 and 42.0

°

C.

Note

When you hold down the direction key, the step size will change to 1.0

°

C

until the key is released.

5. Push  to set the temperature and go back to the Temperature screen.

How to Simulate Respiration

Respiration variables are set through the special functions. To set respiration:
1. Push .
2. Push , , , or  to highlight Respiration.
3. Push . The Respiration screen in Figure 27, shows in the display.

glh013.bmp

Figure 27. Respiration Screen

You can choose between a normal or ventilated respiration waveform and change the

respiration rate. The respiration signal can be set to the left arm (LA) or left leg (LL)

ECG lead. The baseline impedance between the leads and the amplitude of impedance

variation (respiration amplitude) are set though the front panel as well. To set respiration

between normal and ventilated:
1. Push  or  to highlight the Wave value.
2. Push .
3. Push  or  to highlight Normal or Ventilated.
4. Push  to set the wave and go back to the Respiration screen.
To set the respiration rate:
1. Push  or  to highlight the Rate value.
2. Push .
3. Push  or  to increase or decrease the rate of respiration in 1 brpm steps between

10 and 150 brpm.

4. Push  to set the respiration rate and go back to the Respiration screen.

Advertising