Fluke Biomedical ProSim 6 User Manual

Page 51

Advertising
background image

Vital Signs Simulator

How to Simulate Cardiac Output

39

Blood temperat

u

re (BT) connector

Injectate temperat

u

re (IT) thermistor connector

Injectate temperat

u

re (IT) 10-t

u

rn potentiometer

CI-3 mod

u

le/ca

b

le

2

1

4

3

Pin1

Pin3

Crimp t

ub

e

Thermistor end of ca

b

le

ASSEMBLY of the GE

N

ERAL PURPOSE CO

NN

ECTOR

V

ie

w

ed from

the top of

the C1-3

PI

N

1

FU

N

CTIO

N

I

N

JECTATE THERMISTOR CO

NN

ECTIO

N

A

Injectate temperat

u

re (IT) thermistor connector

I

N

JECTATE THERMISTOR CO

NN

ECTIO

N

B

2

N

O CO

NN

ECTIO

N

4

N

O CO

NN

ECTIO

N

3

Solder

w

ires to pins 1 and 3

eic224.eps

Figure 28. Cardiac Output Injectate Cable Modification

Note

Fluke Biomedical offers optional adapter cables to connect the Product to
specified brands of cardiac output measurement devices.

•

To examine Hewlett Packard Merlin systems, a cardiac output adapter

and a temperature adapter are necessary.

•

A special adapter is necessary for Gould/SpectraMed Models SP1445

and SP1465 cardiac output devices.

Contact Fluke Biomedical for availability, price, and part number.

For cardiac output simulation, use the supplied CI-3 module/cable to connect the Product

to the DUT. (see Figure 29). If necessary, use the general-purpose connector.