Fluke Biomedical ProSim 6 User Manual

Page 51

Advertising
background image

Vital Signs Simulator

How to Simulate Cardiac Output

39

Blood temperat

u

re (BT) connector

Injectate temperat

u

re (IT) thermistor connector

Injectate temperat

u

re (IT) 10-t

u

rn potentiometer

CI-3 mod

u

le/ca

b

le

2

1

4

3

Pin1

Pin3

Crimp t

ub

e

Thermistor end of ca

b

le

ASSEMBLY of the GE

N

ERAL PURPOSE CO

NN

ECTOR

V

ie

w

ed from

the top of

the C1-3

PI

N

1

FU

N

CTIO

N

I

N

JECTATE THERMISTOR CO

NN

ECTIO

N

A

Injectate temperat

u

re (IT) thermistor connector

I

N

JECTATE THERMISTOR CO

NN

ECTIO

N

B

2

N

O CO

NN

ECTIO

N

4

N

O CO

NN

ECTIO

N

3

Solder

w

ires to pins 1 and 3

eic224.eps

Figure 28. Cardiac Output Injectate Cable Modification

Note

Fluke Biomedical offers optional adapter cables to connect the Product to
specified brands of cardiac output measurement devices.

To examine Hewlett Packard Merlin systems, a cardiac output adapter

and a temperature adapter are necessary.

A special adapter is necessary for Gould/SpectraMed Models SP1445

and SP1465 cardiac output devices.

Contact Fluke Biomedical for availability, price, and part number.

For cardiac output simulation, use the supplied CI-3 module/cable to connect the Product

to the DUT. (see Figure 29). If necessary, use the general-purpose connector.