How to Set the Non-Invasive Blood Pressure Variables

For non-invasive blood pressure tests, connect the Product to the BP cuff and monitor as

shown in Figure 31.

ProSim 8

Patient

Monitor

Mandrel

Dual hose system: connect Cuff Adapter to
hose marked “Sense”. If both hoses are
unmarked, connect Cuff Adapter to either
hose.

Blood Pressure Cuff

Wraps around mandrel.

Must be connected closer
to the cuff than monitor.

glh031.eps

Figure 31. Non-Invasive Blood Pressure Test Connections

There are two blood pressure cuff mandrels that are used with the Product: Adult and

Neonatal. Figure 32 shows the parts of an adult mandrel and how to assemble it for

different sizes.

Large Adult

Use 2 end blocks and
3 spacer blocks

Use 2 end blocks and
2 spacer blocks

Use 2 end blocks and
1 spacer blocks

Use 2 end blocks and
no spacer blocks

Adult Cuff Mandrel Sizes

Adult

Small Adult

Child

fcv011.eps

Figure 32. Adjustments for Adult Blood Pressure Cuff Mandrel