General Specifications

75

Push

Down

Pull Out

glh046.eps

Figure 63. Battery Removal

To put the battery pack into the Product, align the battery pack with the guides on the

Product and push it into the Product until the latch locks.
The ProSim 6/8 battery is not compatible with the ProSim 4.

General Specifications

Temperature

Operating ............................................................ 10

°

C to 40

°

C (50

°

F to 104

°

F)

Storage ............................................................... -20

°

C to +60

°

C (-4

°

F to +140

°

F)

Humidity................................................................. 10 % to 90 % non-condensing
Altitude................................................................... 3,000 meters (9,843 ft)
Size (L x W x H) ..................................................... 30.22 cm x 14.48 cm x 8.64 cm (11.9 in x 5.7 in x 3.4 in)
Display ................................................................... LCD Color Display
Communication

USB Device Upstream Port ................................ Mini-B connector for control by a computer
USB Host Controller Port.................................... Type A, 5 V output, 0.5 A max load. Connector for keyboard, barcode

reader, and printer

Wireless (ProSim8 Only) .................................... IEEE 802.15.4 for control by a computer
USB Device Virtual COM Port Settings

Baud Rate....................................................... 115,200 bps
Data bits.......................................................... 8 data bits
Stop Bits ......................................................... 1 stop bit
Flow Control.................................................... Hardware (RTS/CTS)

Power ..................................................................... Lithium-Ion rechargeable, 7.2 V, 31 Wh battery, 4300 mAh
Battery Charger..................................................... 100 to 240 V, 50/60 Hz input, 15 V/2.0 A output. For best performance,

the battery charger should be connected to a properly grounded ac
receptacle

Battery Life ............................................................ 9 hours (minimum), 100 NIBP cycles typical
Weight .................................................................... 1.81 kg (4 lb)
Safety Standards................................................... EN/IEC 61010-1:2001

Certifications .........................................................

P, ), 

Electromagnetic Compatibility (EMC)................. EN 61326-1:2006