How to view test results – Fluke Biomedical ProSim 6 User Manual

Page 78

Advertising
background image

ProSim 6/8
Users Manual

66

3. Push  or  to set the parameter to the value observed on the UUT.
4. Push  to set the observed value and show the save screen shown below in the

display.

glh049.bmp

Figure 56. Observed Results Screen

Note

If the observed value is incorrect at this point, you can push the Enter
Observed
softkey and change the observed parameter.

5. Test results are saved with a Pass or Fail prefix in their label. Push the Save as

Pass or the Save as Fail softkey. The saving screen in Figure 57 shows in the

display.

glh040.bmp

Figure 57. Saving Screen

The three-digit number is the test result ID. Within the test ID, each saved test results

is assigned a number in sequence. When the save operation completes, the display

goes back to the screen for the function you just saved. In the case of this example,

the display goes back to the temperature screen.

How to View Test Results

To view the test results in memory:
1. Push .
2. Push , , , and  to highlight View Memory.
3. Push .

Note

If no test results data is saved to memory, an error message shows in the
display when you push
.