Pacemaker waveforms – Fluke Biomedical ProSim 6 User Manual

Page 31

Advertising
background image

Vital Signs Simulator

ECG Function

19

Pacemaker Waveforms

The Product can simulate ECG waveforms with a number of artificial-pacemaker

conditions. Table 11 shows the simulations in the waveform list of the Product with a

description of each.

Table 11. Pacemaker Waveforms

Pacemaker Waveform Label

Waveform Description

Atrial 80 bpm

Atrial pacemaker wave at 80 BPM, with a pacer pulse at the start of each
P wave.

Asynchronous 75 bpm

Asynchronous pacemaker wave with continuous ventricular-paced beats
(75 BPM) and no P waves

Demand Freq Sinus

A “demand” pacemaker wave with frequent sinus beats (forty normal
beats followed by twenty ventricular-paced beats, repeated)

Demand Occ Sinus

A “demand” pacemaker wave with occasional sinus beats (twenty normal
beats followed by forty ventricular-paced beats, repeated)

Atr-Vent Sequential

AV-sequential-pacemaker wave with continuous paced beats, each with
an atrial pulse and a P wave followed by a ventricular-paced pulse and
QRS response

Non-Capture Ventricular-paced

beats,

where one out of ten beats has no heart

response.

Non-Function

Continuous pacer pulses at 75 BPM with no heart response.

To select a pacemaker waveform simulation:
1. From the ECG screen, push  or  to highlight the Wave Group value.
2. Push .
3. Push  or  to highlight TV Paced in the wave group list.
4. Push .
5. Push  or  to highlight the Waveform value.
6. Push .
7. Push  or  to highlight a waveform in the list.
8. Push to set the arrhythmia waveform and go back to the ECG: TV Paced

screen.

The amplitude of all TV Paced waveforms can be adjusted. See the How to Set the ECG

Parameters section above. ECG signal artifacts can also be added to each arrhythmia

signal. See the How to Simulate an ECG Artifact section above.
TV-Paced waveforms include the applicable pacer pulses. The pacer pulses can be

changed through the front panel. All but the Atrial 80 bpm waveform have ventricle

pacemaker variables that can be changed. The Atrial 80 bpm has an atrial pacemaker

value. The Atr-Vent Sequential waveform has an atrial pacemaker value and a ventricle

pacemaker value. To change the pacemaker values push the Atrial Settings or Vent.

Settings softkey.
You can change the amplitude, width, and polarity of the pacemaker signal. To change

the amplitude:
1. Push  or  to highlight the Amplitude value.