Fluke Biomedical ProSim 6 User Manual

Page 92

Advertising
background image

ProSim 6/8
Users Manual

80

Synchronization

Normal Sinus heart rates ................................ 30 to 240 BPM

Maximum rate at 1 ml ................................. 240 BPM achievable with pulses up to 1 ml
Maximum rate at 1.25 ml ............................ 180 BPM

Arrhythmias..................................................... Premature atrial contraction (PAC), Premature ventricular contraction

(PVC), atrial fibrillation, and missed beat.

Leak Test

Target Pressure .................................................. 20 to 400 mmHg
Elapsed time ....................................................... 0:30 to 5:00 minutes:seconds in 30 second steps
Range ................................................................. 0 to 200 mmHg/minute
Internal Leak rate................................................ <2 mmHg/min into 500 ml rigid volume

Pressure Relief Test Range ................................. 100 to 400 mmHg

Oximeter SpO2 Optical Emitter and Detector (optional)

% O2

Range ................................................................. 30 to 100 %
Resolution........................................................... 1 %
Accuracy

With oximeter manufacturer’s R-curve

Saturation within UUT specific range..........

±

(1 count + specified accuracy of the UUT)

Saturation outside UUT specific range ....... monotonic with unspecified accuracy

With Fluke Biomedical R-curves

91 to 100 %.................................................

±

(3 counts + specified accuracy of the UUT)

81 to 90 %...................................................

±

(5 counts + specified accuracy of the UUT)

71 to 80 %...................................................

±

(7 counts + specified accuracy of the UUT)

Below 71 %................................................. monotonic with unspecified accuracy


Heart Rate

Range ................................................................. 30 to 300 BPM in 1 BPM steps. Oximeter SpO2 Optical Emitter and

Detector is synchronized with ECG rate delayed by 150 ms.

Accuracy ............................................................. ±1 % of setting

Transmission (Ratio of detector current to LED current, expressed in parts per million (ppm))

Range ................................................................. 0 to 300.00 ppm
Resolution........................................................... 0.01 ppm
Accuracy ............................................................. +50 %/-30 % for compatible monitors, unspecified for others. Selected

by finger size and color: Dark, thick finger, medium finger, light, thin
finger, neonatal foot.

Pulse Amplitude

Range ................................................................. 0 to 20.00 %
Resolution........................................................... 0.01 %

Artifact

Respiration

Range ............................................................. 0 to 5 % of transmission
Resolution....................................................... 1 %
Rate ................................................................ all ProSim respiration simulation settings

Ambient Light

Range ............................................................. 0 to 5X transmitted light
Resolution....................................................... 1X
Frequency....................................................... DC, 50 Hz, 60 Hz, and 1 to 10 kHz in 1 kHz steps

Masimo Rainbow Technology ............................. Masimo Rainbow technology with an optional adapter supplied by

Masimo that allows the ProSim two wavelength to test the Rainbow
multiple wavelength system

Compatible Manufacturer Products

With manufacturer R-curve................................. Nellcor, Masimo, Nonin, and Nihon Kohden
With Fluke Biomedical R-curve........................... Mindray, GE-Ohmeda, Philips/HP, and BCI