Ecg performance tests, How to set a performance wave – Fluke Biomedical ProSim 6 User Manual

Page 33

Advertising
background image

Vital Signs Simulator

ECG Function

21

ECG Performance Tests

As well as physiological waveforms, the Product can supply signals to measure the

performance of an ECG monitor. A set of performance waveforms are used to measure

the frequency response (high and low), sensitivity, gain drift, internal calibration, stylus

damping, paper speed, linearity, and sweep speed of an ECG monitor. Three more

Product functions are used to measure R wave detection, QRS detection, and tall T wave

rejection of an ECG monitor.

How to Set a Performance Wave

The waveforms in the performance wave group are sine, square, triangle, and pulse.

Note

When a performance wave is set on the Product, outputs for respiration,
blood pressure, and temperature are disabled.

The four waveforms used for ECG performance tests are, sine, square, pulse, and

triangle. The rate and amplitude of these waveforms are adjustable to preconfigured

values. To output a performance wave:
1. Push  to show the ECG screen in the display.
2. Push  or  to highlight the Wave Group value.
3. Push .
4. Push  or  to highlight the Performance value in the wave group list.
5. Push  to show the ECG: Performance screen in Figure 10 in the display.

glh019.bmp

Figure 10. Performance Wave Screen

To change the waveform:
1. Push  or  to highlight the Waveform value.
2. Push .
3. Push  or  to highlight Square, Sine, Triangle, or Pulse in the list.
4. Push  to set the performance waveform and go back to the ECG:

Performance screen.

All waveforms have the same two variables: Rate and Amplitude.
To change the rate:
1. Push  or  to highlight the Rate value.
2. Push .
3. Push  or  to highlight a value in the rate list. Table 12 lists the rates for each type

of wave.

Advertising
This manual is related to the following products: