Vital Signs Simulator

ECG Function

11

All the vital sign simulation parameters are set to the values contained in the pre-defined

simulation.
You can customize the seven factory simulations and seven custom simulations through

the Ansur functionality.

ECG Function

The Product simulates normal heart signals (ECG) as well as heart signals for a variety of

arrhythmias. Heart rate (beats per minute), signal amplitude, and ST segment elevation

are all controlled by the Product through the user interface. Artifacts can also be

simulated.
To measure the ECG performance of a monitor, connect the Product to the monitor as

shown in Figure 6. A maximum of ten ECG leads can be connected to Product.

ProSim 8

Patient

Monitor

ECG

Cable

glh030.eps

Figure 6. ECG Connections

How to Set the ECG Parameters

Whenever the Product is turned on, the display shows the defaults for heart rate

(60 BPM) with a patient type of Adult. Although not shown on the initial display, the

signal amplitude is 1 mV. To set any of the ECG parameters, push . The ECG

screen shown in Figure 7 shows in the display.

glh004.bmp

Figure 7. ECG Screen

Advertising