Vital Signs Simulator

Invasive Blood Pressure Simulation and Tests

31

Note

The artifact values in the list can be in percent (%) or in mmHg. These units
are set by the chamber value.

4. Push  to set the artifact value and go back to the IBP screen.
The IBP screen reappears in the display with the new artifact selection.

How to Simulate Invasive Blood Pressure Tests

The Product can simulate pressures that occur during a Swan-Ganz or Cardiac

Catheterization procedure. Figure 20 shows a monitor connected to the two IBP channel

jacks on the Product.

ProSim 8

Patient

Monitor

IBP Cables

glh028.eps

Figure 20. Invasive Blood Pressure Connections

How to Simulate a Swan-Ganz Procedure

To simulate IBP during a Swan-Ganz procedure:
1. Push the Tests softkey in the IBP screen.
2. Push  or  to highlight Swan-Ganz – Channel 1 or Swan-Ganz –

Channel 2.

3. Push . The details and steps in the procedure shown in Figure 21 are shown in

the display.

Advertising
This manual is related to the following products: