How to do a pressure source test – Fluke Biomedical ProSim 6 User Manual

Page 61

Advertising
background image

Vital Signs Simulator

Non-Invasive Blood Pressure Simulation and Tests

49

If there is no drop in pressure and the pressure climbs to the target pressure, the pump

stops and Not Tripped shows in the display.

Note

Some NIPB monitors do not let you access a “Service” mode. If you can
not close the vent valve, the system can not be pressurized by an external
pump. It is possible to start a blood pressure measurement with the monitor
(this closes the valve), then start the Pressure Relief tests, so that two
pumps inflate the system. The results can change, but the monitor usually
opens a relief valve at some high pressure.

glh008.bmp

Figure 41. Pressure Relief Valve Test Results Screen

See the How to Save Test Results section to learn more on how to save your test results

data.

How to Do a Pressure Source Test

The pressure source test is used to pressurize a pneumatic system while it measures the

pressure. This can be used for static calibration of non-invasive blood pressure

measurement systems, sphygmomanometer checks, and other devices that measure

pressure.
To do a pressure source test:
1. Connect the pressure port to pressure system as shown in Figure 42.

Advertising
This manual is related to the following products: