Vital Signs Simulator

Non-Invasive Blood Pressure Simulation and Tests

43

Figure 33 shows the neonatal blood pressure cuff mandrel.

Large

Medi

u

m

Small

fcv012.eps

Figure 33. Neonatal Blood Pressure Cuff Mandrel

To set the blood pressure simulation, push  to show the NIBP screen in Figure 34 in

the display.

glh003.bmp

Figure 34. Non-Invasive Blood Pressure Screen

Pressure, heart rate, pulse volume, brand, and wave are set through the front-panel

controls to simulate different patient conditions. Arrhythmia waveforms are also

simulated in the NIBP simulation (set through ECG). To set the Blood Pressure:
1. Push  or  to highlight the Pressure variable.
2. Push .

Systolic and diastolic pressures are set separately.

3. Push  or  to highlight the systolic or diastolic pressure value.
4. Push  or  to increase or decrease the pressure value in 1 mmHg steps.

Note

When you hold down the direction key, the step size will change to
10 mmHg until the key is released.

5. Push  to set the pressure and go back to the NIBP screen.