Pulse Rate Test

You can simulate different patient conditions while you monitor the effect of different

pulse rates on the SpO2 measurement. To do a pulse rate test:
1. Set the pulse rate alarm limit value(s) on the oximeter.
2. From the SpO2 screen, push  or  to highlight the Heart Rate value.
3. Push .
4. Push  to increase the heart rate until the alarm sounds.
5. Push  to decrease the heart rate until the alarm sounds.

You can decrease the heart rate to 0 bpm to simulate asystole or no pulse.

Pulse Amplitude Test

The peak-to-peak amplitude of the blood pressure wave tested by the Product can be

increased or decreased. You can decrease the amplitude to find where the oximeter fails

to sense a pulse. To do a pulse amplitude test:
1. From the SpO2 screen, push  or  to highlight the Pulse Amplitude value.
2. Push .
3. Push  to decrease the pulse amplitude value until the oximeter no longer shows a

pulse or the oximeter alarm sounds.

Autosequences

Autosequences are a series of steps that change the output of the Product automatically.

For example, to do a temperature test on a monitor, you must change the temperature of

the temperature simulation a number of times for a specified time period. To do this

manually, you can go to the temperature function, push  and then push  or  to

change the temperature and then push  again. After some time period, you do these

same steps for the subsequent temperature change. The Temperature autosequence does

these changes for you automatically. Each step of the temperature autosequence sets the

temperature and after the allotted time period, it does the subsequent step.
Table 14 is a list of autosequences that are built into the Product.

Table 14. Autosequences

Autosequence

Sequence Steps

Run Time

ECG 200 BPM, IBP1 (Arterial, 200/150), IBP2 (PA,
45/25), SpO2 100 %, NIBP 200/150, Respiration 80
brpm and Temperature 42

°

C

01:30

ECG 80 BPM, IBP1 (Arterial, 120/80), IBP2 (PA, 28/15),
SpO2 97 %, NIBP 120/80, Respiration 20 brpm and
Temperature 37

°

C

01:30

ECG 30 BPM, IBP1 (Arterial, 60/30), IBP2 (PA, 20/10),
SpO2 85 %, NIBP 60/30, Respiration 10 brpm and
Temperature 32

°

C

01:30

Monitor Testing

STOP Total

Time

04:30