Vital Signs Simulator

Autosequences

59

Table 14. Autosequences (cont.)

Autosequence

Sequence Steps

Run Time

ECG 80 bpm, IBP1 120/80 (Art), IBP2 120/0 (LV), SpO2
98 %, NIBP 120/80, Respiration 20 brpm, and
Temperature 37

°

C

01:00

ECG 120 bpm, IBP1 250/150 (Art), IBP2 200/0 (LV),
SpO2 85 %, NIBP 200/150, Respiration 60 brpm, and
Temperature 40

°

C

01:00

ECG 30 bpm, IBP1 60/30 (Art), IBP2 60/0 (LV), SpO2
50 %, NIBP 60/30, Respiration 10 brpm, and
Temperature 34

°

C

01:00

Medical Training

REPEAT Total

Time

03:00

SpO2 100 %, Pulse Amplitude 20 %, Transmissivity:
normal adult, heart rate 60BPM

00:45

SpO2 95 %, Pulse Amplitude 15 %, Transmissivity:
normal adult, heart rate 60BPM

00:45

SpO2 90 %, Pulse Amplitude 10 %, Transmissivity:
dark/thick adult, heart rate 120BPM

00:45

SpO2 85 %, Pulse Amplitude 7 %, Transmissivity:
dark/thick adult, heart rate 30BPM

00:45

SpO2 84 %, Pulse Amplitude 5 %, Transmissivity:
normal adult, heart rate 60BPM

00:45

SpO2 83 %, Pulse Amplitude 4 %, Transmissivity:
normal adult, heart rate 60BPM

00:45

SpO2 82 %, Pulse Amplitude 3 %, Transmissivity:
dark/thick adult, heart rate 120BPM

00:45

SpO2 81 %, Pulse Amplitude 2 %, Transmissivity:
dark/thick adult, heart rate 30BPM

00:45

SpO2 80 %, Pulse Amplitude 1 %, Transmissivity:
normal adult, heart rate 60BPM

00:45

SpO2 79 %, Pulse Amplitude 0.5 %, Transmissivity:
normal adult, heart rate 60BPM

00.45

Oximeter Testing

STOP

Total time 07:30

NSR (Adult) 80 BPM

00:45

Premature PVC1 Left Vent

00:30

Ventricular Mono VTach

00:30

Ventricular Ventricular Fibrillation

00:30

Ventricular Asystole

00:25

Cardiac Failure

STOP Total

Time

02:30