How to view the ecg waveform – Fluke Biomedical ProSim 6 User Manual

Page 25

Advertising
background image

Vital Signs Simulator

ECG Function

13

3. Push  or  to adjust the deviation. Each push of a key moves the deviation 0.05 or

0.1 mV in the direction of the key pushed. The deviation range is

±

0.80 mV

Note

When the ST Deviation value is set to 0.00 mV, the value in ECG screen
will show as Off.

4. Push  to set the deviation and go back to the ECG screen.
To set the ECG signal amplitude:
1. Push  or  to move the highlight in the display to the Amplitude value.
2. Push .
3. Push  or  to adjust the amplitude. Each push of a key increases or decreases the

amplitude 0.05 mV between 0.05 and 0.50 mV and 0.25 mV above 0.50 mV to

5.00 mV in the direction of the key pushed.

4. Push  to set the signal amplitude and go back to the ECG screen.
Table 9 shows the percentage of the signal amplitude value that is put on each ECG lead.

Table 9. ECG Lead Amplitudes

Waveform

I II III V1 V2 V3 V4 V5 V6

Performance

[1]

70 %

100 %

30 %

100 %

100 %

100 %

100 % 100 % 100 %

Normal Sinus

70 %

100 %

30 %

24 %

48 %

100 %

120 % 112 %

80 %

Normal Sinus,
Horizontal Axis

100 %

30 %

-70 %

20 %

40 %

70 %

100 %

80 %

50 %

Normal Sinus,
Vertical Axis

30 %

130 %

100 %

20 %

30 %

40 %

60 %

80 %

100 %

[1]

Includes R-Wave Detection, QRS Detection, and Tall T-Wave Rejection waves.

How to View the ECG Waveform

As you adjust the ECG signal, you can see the signal in the display. To see the ECG

signal, from the ECG screen, push the Graph softkey. The graph screen in Figure 8

shows in the display.

glh023.bmp

Figure 8. ECG Graph Screen

The ECG signal is shown in the display in real time.
Push the Back softkey to go back to the ECG screen.

Advertising
This manual is related to the following products: