Fluke Biomedical ProSim 6 User Manual

Page 22

Advertising
background image

ProSim 6/8
Users Manual

10

Table 7. Pre-Defined Patient Simulations (cont.)

Simulation Name

Parameter

Pre-Set Value

Wave Form

NSR (Adult)

ECG Rate

30 bpm

Respiration Rate

15 brpm

Temperature 35.0

°

C

IBP Channel 1

100/65 mmHg (Art)

IBP Channel 2

25/10 mmHg (PA)

NIBP

100/65 (77) mmHg

Bradycardic

SpO2 85%

Wave Form

Atrial Fibrillation (Coarse)

ECG Rate

No bpm

Respiration Rate

70 brpm

Temperature 34.0

°

C

IBP Channel 1

35/15 mmHg (Art)

IBP Channel 2

35/0 mmHg (LV)

NIBP

35/15 (22) mmHg

Ventricular Fibrillition

SpO2 80

%

Wave Form

Asystole

ECG Rate

0 bpm

Respiration Rate

0 brpm

Temperature 33.0

°

C

IBP Channel 1

0 mmHg

IBP Channel 2

0 mmHg

NIBP 0/0

mmHg

Asystole

SpO2 50

%

After you set one of the pre-defined simulations, you can change the parameters of one or

all of the vital signs. Changes are not kept in the Product. The pre-defined simulation

parameters can not be changed through the front panel.
There are seven custom pre-defined simulations in the Product. To set a custom

simulation:
1. From the Home screen, push the More softkey.
2. Push the Custom softkey.

Four custom simulation names show above the first four softkeys. The names are

Custom1, Custom2, Custom3, and Custom4. To show the next three names,

push the More softkey.

3. Push the softkey for a custom pre-defined simulation.