Detailed specifications, Normal-sinus-rhythm waveform, Pacemaker waveform – Fluke Biomedical ProSim 6 User Manual

Page 88: Arrhythmia

Advertising
background image

ProSim 6/8
Users Manual

76

Detailed Specifications

Normal-Sinus-Rhythm Waveform

ECG Reference...................................................... The ECG amplitudes specified are for Lead II (calibration), from the

baseline to the peak of the R wave. All other leads are proportional.

Normal Sinus Rhythm .......................................... 12-lead configuration with independent outputs referenced to right leg

(RL). Output to 10 Universal ECG Jacks, color-coded to AHA and IEC
Standards.

High-Level Output................................................. 0.5 V/mV

±

5 % of the ECG amplitude setting available on a BNC

connector.

Amplitude .............................................................. 0.05 mV to 0.5 mV (0.05 mV steps); 0.5 mV to 5.0 mV (0.25 mV

steps). Other leads are proportional to Lead II (reference lead) in
percentage per:

Lead I .................................................................. 70
Lead II ................................................................. 100
Lead III ................................................................ 30
Lead V1 .............................................................. 24
Lead V2 .............................................................. 48
Lead V3 .............................................................. 100
Lead V4 .............................................................. 120
Lead V5 .............................................................. 112
Lead V6 .............................................................. 80

Amplitude Accuracy .............................................

±

(2 % of setting + 0.05 mV)

ECG Rate

ProSim 8 ............................................................. 10 to 360 BPM in 1 BPM steps
ProSim 6 ............................................................. 30 to 360 BPM in 10 BPM steps

Rate Accuracy .......................................................

±

1 % of setting

ECG Waveform Selection..................................... Adult (80 ms) or pediatric (40 ms) QRS duration
ST-Segment Elevation .......................................... Adult mode only. -0.8 mV to +0.8 mV (0.1 mV steps) Additional steps:

+0.05 mV and -0.05 mV

Power-On Default.................................................. 60 BPM, 1.0 mV, adult QRS and ST-segment elevation of 0 mV

Pacemaker Waveform

Pacer-Pulse

Amplitude .......................................................... 0 (off),

±

2,

±

4,

±

6,

±

8,

±

10,

±

12,

±

14,

±

16,

±

18,

±

20,

±

50,

±

100,

±

200,

±

500, and

±

700 mV for lead II (reference lead)

Accuracy

Reference lead II ............................................

±

(5 % setting + 0.2 mV)

All other leads .................................................

±

(10 % setting + 0.4 mV)

Pacer-Pulse Width ................................................ 0.1, 0.2, 0.5, 1.0, and 2.0 ms

±

5 %

Paced Arrhythmias ............................................... Atrial 80 BPM

Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
Atrio-Ventricular sequential
Noncapture (one time)
Nonfunction

Power-On Default.................................................. Amplitude 10 mV, width 1.0 ms, atrial waveform

Arrhythmia

Baseline NSR......................................................... 80 BPM
PVC Focus ............................................................. Left focus, standard timing (except where specified)
Supraventricular Arrhythmia

ProSim 8 ............................................................. Atrial fibrillation (coarse or fine); atrial flutter; sinus arrhythmia; missed

beat (one time); atrial tachycardia; paroxysmal atrial tachycardia; nodal
rhythm; and supraventricular tachycardia.

ProSim 6 ............................................................. Atrial fibrillation (coarse or fine) and sinus arrhythmia

Premature Arrhythmia