Fluke Biomedical ProSim 6 User Manual

Page 65

Advertising
background image

Vital Signs Simulator

Oximeter SpO2 Optical Emitter and Detector

53

Connect the SpO2 artificial finger to the SpO2 jack on the front panel as shown in

Figure 47.

ProSim 8

SpO

2

Cable

Pulse

Oximeter

Sensor

Patient

Monitor

ProSim

Oximeter SpO

2

Optical

Emitter and Detector

glh029.eps

Figure 47. Oximeter SpO2 Optical Emitter and Detector Connections

Note

When you put the oximeter sensor on the artificial finger, make sure the red
LEDs (light emitting diodes) are on the bottom.

Put the SpO2 sensor on the artificial finger as shown in Figure 48. Place the sensor with

the LEDs on the bottom of the artificial finger. While you put the sensor on the artifical

finger, monitor the signal indicator along the bottom of the Product display. Adjust the

sensor on the finger for maximum signal strength.