Fluke Biomedical ProSim 6 User Manual

Page 24

Advertising
background image

ProSim 6/8
Users Manual

12

To change the ECG waveform:
1. Push  or  to move the highlight to the Wave Group value.
2. Push .
3. Push  or  to move the highlight to a waveform group name. Table 8 is a list of

the wave groups available in the Product.

Table 8. ECG Wave Groups

Wave group

NSR (Adult)

Adult Normal Sinus Rhythm

NSR (Pediatric)

Child Normal Sinus Rhythm

Performance

R-Wave Detection (ProSim 8 only)

QRS Detection (ProSim 8 only)

Tall T Wave Rejection (ProSim 8 only)

Special waveforms for monitor tests. See the ECG
Tests section.

Supraventricular

Premature

Ventricular

Conduction

TV Paced

ACLS (ProSim 8 only)

Arrhythmia waveforms. See the How to Simulate
Arrhythmias section.


4. When the waveform group value is highlighted, push .
To change the heart rate:
1. Push  or  to move the highlight in the display to the Heart Rate value.
2. Push .
3. Push  or  to adjust the heart rate. Each push of a key moves the heart rate one

beat in the direction of the key pushed. The ProSim 8 simulates a heart rate of 10 to

360 beats/minute (BPM) set in 1 BPM intervals. The ProSim 6 simulates a heart rate

of 30 to 360 beats/minute (BPM) set in 10 BPM intervals.

4. Push  to set the heart rate and go back to the ECG screen.
To change the axis:
1. Push  or  to move the highlight to the Axis value.
2. Push .
3. Push  or  to highlight Intermediate, Horizontal, or Vertical.
4. Push .
To set the ST Deviation:
1. Push  or  to move the highlight to the ST Deviation value.
2. Push .