Ecg-performance-testing – Fluke Biomedical ProSim 6 User Manual

Page 89

Advertising
background image

Vital Signs Simulator

Detailed Specifications

77

ProSim 8 ............................................................. Premature atrial contraction (PAC); premature nodal contraction

(PNC); PVC1 left ventricular; PVC1 left ventricular, early; PVC1 left
ventricular, R on T; PVC2 right ventricular; PVC2 right ventricular,
early; PVC2 right ventricular, R on T; and multifocal PVCs

ProSim 6 ............................................................. Premature atrial contraction (PAC) and multifocal PVCs

Ventricular Arrhythmia

ProSim 8 ............................................................. PVCs 6, 12, or 24 per minute; frequent multifocal PVCs; bigeminy;

trigeminy; multiple PVCs (one-time run of 2, 5, or 11 PVCs); mono-
ventricular tachycardia (120 to 300 BPM in 5 BPM steps); poly-
ventricular tachycardia (5 types); ventricular fibrillation (coarse or fine);
and Asystole

ProSim 6 ............................................................. Bigeminy; mono-ventricular tachycardia (120 to 300 BPM in 5 BPM

steps); poly-ventricular tachycardia (5 types); ventricular fibrillation
(coarse or fine); and Asystole

Conduction Defect

ProSim 8 ............................................................. First-, second-, or third-degree heart block; and right- or left-bundle-

branch block

ProSim 6 ............................................................. Second-degree heart block, type 1, and right-bundle-branch block

Advanced Cardiac Life Support (ProSim 8 only)

Shockable Pulseless Arrest Rhythms................. Ventricular fibrillation (coarse), ventricular fibrillation (fine), unstable

polymorphic ventricular tachycardia

Non Shockable Pulseless Arrest Rhythms ......... Asystole
Symptomatic Bradycardia................................... Sinus Bradycardia (<60 BPM)

2

nd

Degree AV Block, Mobitz Type I

2

nd

Degree AV Block, Mobitz Type II

Complete/3

rd

Degree AV Block

Right Bundle Branch Block
Left Bundle Branch Block

Symptomatic Tachycardia

Regular Narrow-complex Tachycardias (QRS <0.12 seconds)

Sinus Tachycardia ...................................... >150 BPM
Supraventircular Tachycardia ..................... SVT

Regular Wide-complex Tachycardias (QRS

0.12 seconds)

Sinus Tachycardia ...................................... >150 BPM
Supraventircular Tachycardia ..................... SVT with aberrancy

Irregular Tachycardia...................................... Atrial Fibrillation (Coarse and fine), Atrial Flutter, unstable

monomorphic ventricular tachycardia (120 to 300 BPM), Torsade De
Pointes/Polymorphic ventricular tachycardia (long QT interval)

ECG-Performance-Testing

Amplitude (peak-to-peak)..................................... 0.05 to 0.5 mV (0.05 mV steps)

0.5 to 5.0 mV (0.25 mV steps)
Other leads are proportional to Lead II (reference lead) in percentage
per:

Lead I.................................................................. 70

Lead II ................................................................. 100
Lead III ................................................................ 30
Lead V1 through V6............................................ 100

Pulse Wave ............................................................ 30, 60 BPM, with 60 ms pulse width
Square Wave ......................................................... 0.125, 2.0, 2.5 Hz
Triangle Wave........................................................ 0.125, 2.0, 2.5 Hz
Sine Wave .............................................................. 0.05, 0.5, 1, 2, 5, 10, 25, 30, 40, 50, 60, 100, and 150 Hz
R-wave Detection (ProSim 8 only)

Waveform............................................................ Triangular pulse
Rate .................................................................... 30, 60, 80, 120, 200, and 250 BPM
Width................................................................... 8 to 20 ms in 2 ms steps, and 20 to 200 ms in 10 ms steps
Width Accuracy...................................................

±

(1 % of setting + 1 ms)

QRS Detection (ProSim 8 only)

Widths................................................................. 8 to 20 ms in 2 ms steps and 20 to 200 ms in 10 ms steps
Width Accuracy...................................................

±

(1 % of setting + 1 ms)

Advertising
This manual is related to the following products: