How to clean the product, Battery maintenance, How to charge the battery – Fluke Biomedical ProSim 6 User Manual

Page 85

Advertising
background image

Vital Signs Simulator

Maintenance

73

Do not put battery cells and battery packs near heat or fire.

Do not put in sunlight.

Do not disassemble or crush battery cells and battery

packs.

To prevent possible electrical shock, fire, or personal injury:

Remove the input signals before you clean the Product.

Use only specified replacement parts.

Have an approved technician repair the Product.

How to Clean the Product

W Caution

Do not pour fluid onto the Product surface; fluid seepage into

the electrical circuitry may cause the Product to fail.

W Caution

Do not use spray cleaners on the Product; such action may

force the cleaning fluid into the Product and damage electronic

components.

Clean the Product occasionally with a damp cloth and mild detergent. Try to prevent the

entrance of liquids.
Clean the adapter cables with the same precautions. Examine them for damage and

deterioration of the insulation. Examine the connections for integrity. Keep the transducer

adapter clean and dry.

Battery Maintenance

For peak battery performance, charge the Product to maximum charge once a month. If

the Product is not to be used for more than a month, keep it connected to the charger.

Note

To get the specified performance, use the specified battery charger that
comes with this Product.

When the battery gets low, a low battery message shows in the display.
When the battery discharges to 3 % of full charge, a different message shows and the

NIBP function is disabled.

How to Charge the Battery

The battery charge level is shown in the upper-right corner of the display when the

battery pack is installed in the Product. If the Product is plugged in, … shows in the

upper right corner of the display. When the battery charges, the status is updated in

Battery setting under . With the AC/DC power supply removed from the Product,

the battery icon shows the charge level.
The battery can be charged while it is in or out of the Product. The charge rate is slower

when the Product is energized and the battery charger is on. To charge the battery:
1. As shown in Figure 62, connect the AC/DC power supply to the power connector on

the battery pack.

2. Connect the AC/DC power supply to a power source.

Advertising