Fluke Biomedical ProSim 6 User Manual

Page 42

Advertising
background image

ProSim 6/8
Users Manual

30

4. Push  to set the chamber and go back to the IBP screen.
You must first zero the patient monitor when you connect to a new IBP transducer or

patient simulator. To set a channel to 0 mmHg and Static chamber, push the Zero

Channel 1 or Zero Channel 2 softkey. After you zero the product, do the zero

function on the patient monitor.
To set the pressure:
1. Push  or  to highlight the Pressure value on channel 1 or channel 2.
2. Push .

Systolic and diastolic pressures are set separately.

Note

If the chamber is set to static, only one pressure is set.

3. Push  or  to highlight the systolic or diastolic pressure value.
4. Push  or  to increase or decrease the pressure value in 1 mmHg steps. The range

is -10 to +300 mmHg.

Note

When you hold down the direction key, the step size will change to
10 mmHg until the key is released.

5. Push  to set the pressure and go back to the IBP screen.
When the Chamber parameter is set to a value other than Static, the artifact parameter is

added to the display as shown in Figure 19.

glh051.eps

Figure 19. IBP with Chambers Screen

To set the artifact:

Note

Artifact value is not available when the chamber value is set to static.

1. Push  or  to highlight the Artifact value on channel 1 or channel 2.
2. Push .
3. Push  or  to highlight 5% or 10% if the chamber parameter is set to Arterial,

Radial Artery, or Left Ventricle. Highlight 5 mmHg or 10 mmHg for all other

chamber values. Highlight Off to turn off the artifact function