ProSim 6/8
Users Manual

20

2. Push .
3. Push  or  to highlight an amplitude value in the list. Below is a list of the

pacemaker amplitude values.

0 mV

16 mV

2 mV

18 mV

4 mV

20 mV

6 mV

50 mV

8 mV

100 mV

10 mV

200 mV

12 mV

500 mV

14 mV

700 mV


4. Push  to set the amplitude and go back to the TV Paced Settings screen.
To change the width value:
1. Push  or  to highlight the Width value.
2. Push .
3. Push  or  to highlight 0.1, 0.2, 0.5, 1.0, 1.5, or 2.0 mV in the list.
4. Push  to set the width and go back to the TV Paced Settings screen.
To change the polarity:
1. Push  or  to highlight the Polarity value.
2. Push .
3. Push  or  to highlight + or - in the list.
4. Push  to set the polarity and go back to the TV Paced Settings screen.
Push the Back softkey to go back to the ECG: TV Paced screen.

Advance Cardiac Life Support (ACLS) Waveforms (ProSim 8 only)

The Product can simulate Advanced Cardiac Life Support (ACLS) waveforms.
To simulate an ACLS waveform:
1. From the ECG screen, push  or  to highlight the Wave Group value.
2. Push .
3. Push  or  to highlight ACLS in the wave group list.
4. Push .
5. Push  or  to highlight the Waveform value.
6. Push .
7. Push  or  to highlight a waveform in the list.
8. Push to set the waveform and go back to the ECG: ACLS screen.
The amplitude of all ACLS waveforms can be adjusted. See the How to Set the ECG

Parameters section above. ECG signal artifacts can also be added to each arrhythmia

signal. See the How to Simulate an ECG Artifact section above.