Ecg artifact, Fetal / maternal-ecg (prosim 8 only), Invasive blood pressure – Fluke Biomedical ProSim 6 User Manual

Page 90

Advertising
background image

ProSim 6/8
Users Manual

78

Rate .................................................................... 30, 60, 80, 120, 200, and 250 BPM
R-Wave up slope ................................................ 0.875 amplitude, 0.4375 x width
R-Wave down slope............................................ Full amplitude, 0.5 x width
S-Wave up slope ................................................ 0.125 amplitude, 0.0625 x width

Tall T-Wave Rejection (ProSim 8 only)

Waveform

QT Interval ...................................................... 350 ms
T-Wave width.................................................. 180 ms
T-Wave Shape................................................ ½ sinewave

Amplitude............................................................ 0 to 150 % reference lead amplitude in 10 % steps
Rate .................................................................... 80 BPM

Rate Accuracy .......................................................

±

1 % of setting

Amplitude Accuracy .............................................

±

(2 % of setting + 0.05 mV)

ECG Artifact

Type........................................................................ 50 Hz, 60 Hz, muscular, baseline wander, respiration
Size......................................................................... 25, 50, 100 % of the normal sinus R-Wave for each lead
Lead Select ............................................................ All, RA, LL, LA, V1, V2, V3, V4, V5, V6

Fetal / Maternal-ECG (ProSim 8 only)

Fetal Heart Rate (Fixed)........................................ 60 to 240 BPM in 1 BPM steps
Fetal Heart Rate (IUP) ........................................... 140 BPM at beginning, then varies with pressure
Intrauterine-Pressure Waveforms ....................... Early deceleration, late deceleration, and acceleration
Wave Duration....................................................... 90 seconds, bell-shaped pressure curve, from 0 to 90 mmHg and

returning to 0

IUP Period.............................................................. 2, 3, or 5 minutes; and manual
Default Settings..................................................... FHR 140 BPM, early deceleration wave, manual

Invasive Blood Pressure

Channels................................................................ 2, each independently settable with identical parameters and are

individually electrically isolated from all other signals

Input/output Impedance ....................................... 300

Ω

±

10 %

Exciter Input Range .............................................. 2.0 to 16.0 V peak
Exciter-Input Frequency Range........................... DC to 5000 Hz
Transducer Sensitivity ......................................... 5 (default) or 40

μ

V/V/mmHg

Pressure Accuracy ...............................................

±

(1 % of setting + 1 mmHg) Accuracy guaranteed for DC excitation

only

Static Pressure...................................................... -10 to +300 mmHg in 1 mmHg steps
Pressure Units....................................................... mmHg or Kpa
Dynamic Waveforms

Types (default pressures) ................................... Arterial (120/80)

Radial artery (120/80)
Left ventricle (120/00)
Right ventricle (25/00)
Pulmonary artery (25/10)
Pulmonary-artery wedge (10/2)
Right atrium (central venous or CVP) (15/10)

Pressure Variability............................................. Systolic and diastolic pressures are independently variable in 1 mmHg

steps.

Swan-Ganz Sequence........................................... Right atrium, right ventrical (RV), pulmonary artery (PA), pulmonary

artery wedge (PAW)

Cardiac Catheterization (ProSim 8 only)

Chambers ........................................................... Aortic, Pulmonary valve, and Mitral valve

Respiration Artifact

Arterial, radial artery, and left ventricle ............... 5 to 10 % multiplication
Other ................................................................... 5 or 10 mmHg

BP Output .............................................................. Circular DIN 5-Pin
Power-On Default.................................................. 0 mmHg