Fluke Biomedical ProSim 6 User Manual

Page 62

Advertising
background image

ProSim 6/8
Users Manual

50

ProSim 8

Mandrel

Blood Pressure Cuff

Blood Pressure

Gauge

Wraps around mandrel.

glh051.eps

Figure 42. Pressure Source Connection

2. From the NIBP screen, push the Tests softkey.
3. Push  or  to highlight Pressure Source.
4. Push .
5. Push  or  to increase or decrease the target pressure between 20 and 400 mmHg

in 1 mmHg steps.

Note

When you hold down the direction key, the step size will change to
10 mmHg until the key is released.

6. Push the Start softkey to begin the test.
The Product pressurizes the pneumatic system to the target pressure. While the Product

pumps air into the pneumatic system, the pressure measurement and a graph of the

pressure is shown in the display. See Figure 43.

glh010.bmp

Figure 43. Pressure Source Pumping Screen